Percussionaire Corporation: Phasitron 5, IPV In-Line Valve, (Intrapulmonary Percussive Ventilation) REF: P5-TEE-20 (pack of 20)/ P5-TEE (individual) with In-Line Valve
Read the official notice at FDA (device) →
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Percussionaire Corporation
- Product
- Phasitron 5, IPV In-Line Valve, (Intrapulmonary Percussive Ventilation) REF: P5-TEE-20 (pack of 20)/ P5-TEE (individual) with In-Line Valve Sentec Instructions for Use
- Reason / hazard
- IFU update; in-line valve, with pressure relief seal may leak when used with ventilator; don't use in neonates/infants, with target tidal volumes less-than50mL or less-than10kg- at high risk of cardiopulmonary/neurological compromise due to unrecognized hypoventilation, updates for continued use in others before design update; leak may cause respiratory acidosis, hypoxemia, or respiratory failure.
- Classification
- Class I
- Status
- Ongoing
- Distribution / sold at
- Worldwide distribution. US nationwide, Russia, Japan, Switzerland, Canada, Qatar, Chile, South Africa, United Kingdom, Israel, Turkey, and UAE
- Recall initiated
- 2025-09-17
- Code info
- UDI-DI: 00849436000723. Lots: 240326, 250116, 240418, 250324, 240610, 250616, 240620, 241121, 230612, 240826, WO04884, 241118, WO06020, 241203, WO05019, 250111, WO04827, WO04756. Instructions for Use P20020 Rev F
- Product quantity
- 59160
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Sandpoint
- State
- ID
- Country
- United States
- Event id
- 97374