GE Medical Systems, LLC: LOGIQ P10 R4.5 HD ultrasound system with system software version R4.5.7., Model Number 5877535

Medical DevicesSource: FDA (device) · Date: 2025-10-22

Read the official notice at FDA (device) →

Share on XFacebookRedditEmail

Get alerts for GE Medical Systems

Guides: Which agency handles which recall (FDA vs USDA vs CPSC vs NHTSA)

The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.

Company / brand
GE Medical Systems, LLC
Product
LOGIQ P10 R4.5 HD ultrasound system with system software version R4.5.7., Model Number 5877535
Reason / hazard
The Ultrasound-Guided Attenuation Parameter (UGAP) measurement data may display inaccurate values representing liver steatosis. This could potentially lead to inappropriate clinical decisions impacting overall care.
Classification
Class II
Status
Ongoing
Distribution / sold at
US Nationwide. Global Distribution.
Recall initiated
2025-09-18
Code info
Model Number 5877535 UDI-DI: 00195278464668, Serial Numbers: LPZ490338 LPZ490347 LPZ490339 LPZ490147 LPZ490292 LPZ490019 LPZ490252 LPZ490348 LPZ490303 LPZ490304 LPZ490305 LPZ490306 LPZ490340 LPZ490350 LPZ490341 LPZ440658 LPZ440673 LPZ440674 LPZ490337 LPZ440641 LPZ440192 LPZ440258 LPZ440694 LPZ440695 LPZ440692 LPZ440632 LPZ440185 LPZ440107 LPZ440168 LPZ440046 LPZ440549 LP9490850
Product quantity
32 units
Voluntary mandated
Voluntary: Firm initiated
City
Waukesha
State
WI
Country
United States
Event id
97726
Free daily digest + "My products" watchlist

Get the daily digest, and tell us what to watch. Everything is optional except your email.

Free allergen alerts:

Categories:

Privacy: we store only what you enter here, use it only to send recall alerts you asked for, never sell or share it, and you can unsubscribe anytime. Double opt-in: we will email you a confirmation link before sending anything; alerts start only after you confirm. Every email has unsubscribe and manage-watchlist links.