American Contract Systems Inc.: Basic Biopsy Tray; Item Number/Description: AKRB26D ROBOTIC, AMGL09AA GENERAL LAPAROSCOPY, ANLC35AD LAP CHOLE PACK - 205956, ANLP25A
Read the official notice at FDA (device) →
Get alerts for American Contract Systems
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- American Contract Systems Inc.
- Product
- Basic Biopsy Tray; Item Number/Description: AKRB26D ROBOTIC, AMGL09AA GENERAL LAPAROSCOPY, ANLC35AD LAP CHOLE PACK - 205956, ANLP25AJ LAPROSCOPY PACK - 205958, FYLA38E LAPCHOLAPPY PACK, IHDA88AH DIVINCI GYN URO, IHLC11AE LAP GENERAL PACK, MHLP45AJ LAPAROSCOPY PACK - 206005, MHRB88AO ROBOTIC PACK - 242561, RCLC51C LAP CHOLE RCH 0246969, SMLC53C LAP CHOLE PACK, UDLC17AE DSC LAP CHOLE PACK - 206062, UIRU84AE ROBOTIC UROLOGY PACK (PS 060571)
- Reason / hazard
- Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- US Nationwide distribution in the states of SD, IA, MN, WA, IL.
- Recall initiated
- 2025-09-04
- Code info
- AKRB26D UDI-DI 191072229366 Lot 7350411 Bag Serial Number 68236939; AMGL09AA UDI-DI 191072207357 Lot 8459111 Bag Serial Numbers 68678361 68678364; ANLC35AD UDI-DI 191072211477 LOT 8021911 Bag Serial Number n/a Lot 8173811 Bag Serial Numbers 17556328 17556374 17556376 17556393; ANLP25AJ UDI-DI 191072236333 Lot 8306311 Bag Serial Numbers 47862596 47862598; FYLA38E UDI-DI 191072203519 Lot 8134111 Bag Serial Number 17488260; IHDA88AH UDI-DI 191072226303 Lot 8096211 Bag serial number N/A Lot 8150511 Bag Serial number 68277361; IHLC11AE UDI-DI 191072226372 Lot 8107911 Bag Serial Number N/A Lot 8192711 Bag Serial Numbers 68195543 68196275; MHLP45AJ UDI-DI 191072236500 Lot 8527411 Bag serial number 68873222; MHRB88AO UDI-DI 191072236531 Lot 8322811 Bag serial number 68596086; RCLC51C UDI-DI 191072235930 Lot 8230311 Bag serial number N/A; SMLC53C UDI-DI 191072223029 Lot 8195611 Bag serial numbers 68198164 68198213 68198222 68198228 68198229; UDLC17AE UDI-DI 191072214072 Lot 8501711 Bag serial number 17744347; UIRU84AE UDI-DI 191072235565 Lot 8319311 Bag serial number 68244060 * If "N/A" for bag serial number, then whole lot is affected
- Product quantity
- 32,433 total
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Grimes
- State
- IA
- Country
- United States
- Event id
- 97659