American Contract Systems Inc.: Medical convenience kits UDPG88W PICC G-TUBE PACK-230209

Medical DevicesSource: FDA (device) · Date: 2025-10-22

Read the official notice at FDA (device) →

Share on XFacebookRedditEmail

Get alerts for American Contract Systems

Guides: Which agency handles which recall (FDA vs USDA vs CPSC vs NHTSA)

The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.

Company / brand
American Contract Systems Inc.
Product
Medical convenience kits UDPG88W PICC G-TUBE PACK-230209
Reason / hazard
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
Classification
Class II
Status
Ongoing
Distribution / sold at
US Nationwide distribution in the states of SD, IA, MN, WA, IL.
Recall initiated
2025-09-04
Code info
UDPG88W UDI-DI 191072216359 lot 7935611 bag serial number N/A * If bag serial number is listed as "N/A", then whole lot is affected
Product quantity
32,433 total
Voluntary mandated
Voluntary: Firm initiated
City
Grimes
State
IA
Country
United States
Event id
97659
Free daily digest + "My products" watchlist

Get the daily digest, and tell us what to watch. Everything is optional except your email.

Free allergen alerts:

Categories:

Privacy: we store only what you enter here, use it only to send recall alerts you asked for, never sell or share it, and you can unsubscribe anytime. Double opt-in: we will email you a confirmation link before sending anything; alerts start only after you confirm. Every email has unsubscribe and manage-watchlist links.