American Contract Systems Inc.: Medical convenience kits IHDC30R D AND C PACK
Read the official notice at FDA (device) →
Get alerts for American Contract Systems
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- American Contract Systems Inc.
- Product
- Medical convenience kits IHDC30R D AND C PACK
- Reason / hazard
- Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- US Nationwide distribution in the states of SD, IA, MN, WA, IL.
- Recall initiated
- 2025-09-04
- Code info
- IHDC30R UDI-DI 191072226310 lot 8096511 bag serial number N/A * If bag serial number is listed as "N/A", then whole lot is affected
- Product quantity
- 32,433 total
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Grimes
- State
- IA
- Country
- United States
- Event id
- 97659
Free daily digest + "My products" watchlist