SEDECAL SA: Model Number 40KWFX, Mobile X-ray system
Read the official notice at FDA (device) →
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- SEDECAL SA
- Product
- Model Number 40KWFX, Mobile X-ray system
- Reason / hazard
- Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- US: CA, IL, and NJ
- Recall initiated
- 2025-06-10
- Code info
- UDI/DI 08436046002166, Serial Numbers: G71453, G70953, G70930, G70537, G70757, G70797, G70275, G71257, G71983, G71975, G71467, G69702, G71440, G71329, G70244, G70955, G71462, G71320, G71451, G71466, G71335, G71446, G65260, G70951, G70814, G70931, G70942, G71991, G70261, G72080, G69697, G71978, G70938, G70247, G70250, G70947, G69706, G78917, G70943, G71343, G71345, G71346, G71258, G70527, G69551, G71442, G71977, G71987, G72083, G70242, G72111, G70256, G70806, G71339, G69547.
- Product quantity
- 55 units
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Algete
- Country
- Spain
- Event id
- 97075