GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING: Elscint Cardial; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
Read the official notice at FDA (device) →
Get alerts for Ge Medical Systems Israel, Functional Imaging
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING
- Product
- Elscint Cardial; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
- Reason / hazard
- GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- Worldwide
- Recall initiated
- 2025-09-12
- Code info
- all serial numbers in distribution
- Product quantity
- 11 units
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Haifa
- Country
- Israel
- Event id
- 97647
Free daily digest + "My products" watchlist