Vantive US Healthcare LLC: Sharesource Connectivity Platform for Use with Homechoice Claria Product Code 5CGM01 software versions 7.13.3 (China), 8.13.2 (APAC and
Read the official notice at FDA (device) →
Get alerts for Vantive US Healthcare
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Vantive US Healthcare LLC
- Product
- Sharesource Connectivity Platform for Use with Homechoice Claria Product Code 5CGM01 software versions 7.13.3 (China), 8.13.2 (APAC and Korea), and 8.13.4 (US, EMEA, and LATAM)
- Reason / hazard
- Vantive has identified a software defect within the Sharesource Claria software, which may cause a patient s updated prescription program settings to not be properly saved on the actual cycler that the patient is using to perform their therapy. As a result, the clinician may think that the patient is receiving the correct prescribed therapy because of what is erroneously displayed on the clinician's remote Sharesource application, when in fact the cycler that is with the patient is actually delivering an outdated and incorrect therapy.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- Worldwide distribution - US Nationwide and the countries of Australia, Hong Kong, Korea, New Zealand, Taiwan, Thailand, France, Germany, Ireland, Netherlands, Italy, Spain, Turkey, United Kingdom, Chile, and Colombia.
- Recall initiated
- 2025-08-28
- Code info
- UDI-DI 15413765545723 software versions 7.13.3 (China), 8.13.2 (APAC and Korea), and 8.13.4 (US, EMEA, and LATAM)
- Product quantity
- 1
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Deerfield
- State
- IL
- Country
- United States
- Event id
- 97527