Medline Industries, LP: Convenience kits; DYNJ901047R ORTHO SPINE FUMICH DYNJ902308K TOTAL JOINT DYNJ909111F DR SALMA'S SPINE DYNJ909117G PATTERSON TOTAL JOINT
Read the official notice at FDA (device) →
Get alerts for Medline Industries
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Medline Industries, LP
- Product
- Convenience kits; DYNJ901047R ORTHO SPINE FUMICH DYNJ902308K TOTAL JOINT DYNJ909111F DR SALMA'S SPINE DYNJ909117G PATTERSON TOTAL JOINT DYNJ909113F KYPHOPLASTY
- Reason / hazard
- Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.
- Classification
- Class I
- Status
- Ongoing
- Distribution / sold at
- International distribution to Mexico and Panama. US nationwide distribution to the following: AK AL AR AZ CA CO CT FL GA IA IL IN KS KY MA ME MI MN MO MS MT NC NJ NY OH OR PA PR RI SC TN TX WA WI;
- Recall initiated
- 2025-10-10
- Code info
- DR SALMA'S SPINE DYNJ909111F UDI-DI 40198459130794 (case), 10198459130793 (ea) Lots 25FBE036 25GBV759; ORTHO SPINE FUMICH DYNJ901047R UDI-DI 40198459130466 (case), 10198459130465 (ea) Lots 25FBR230 25GBU693; PATTERSON TOTAL JOINT DYNJ909117G UDI-DI 40198459130480 (case), 10198459130489 (ea) Lots 25FBO307 25GBN625; TOTAL JOINT DYNJ902308K UDI-DI 40198459130596 (case), 10198459130595 (ea) Lots 25FBB321 25GBU688 KYPHOPLASTY DYNJ909113F UDI-DI 40198459130459 (case), 10198459130458 (ea) Lot 25IBM788
- Product quantity
- 96,406 total units
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Northfield
- State
- IL
- Country
- United States
- Event id
- 97882