Philips Respironics, Inc.: SimplyGo Mini, Standard Battery Kit; Product code: 1116816; The affected battery kits are sold as accessories for the SimplyGo Mini: Si
Read the official notice at FDA (device) →
Get alerts for Philips Respironics
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Philips Respironics, Inc.
- Product
- SimplyGo Mini, Standard Battery Kit; Product code: 1116816; The affected battery kits are sold as accessories for the SimplyGo Mini: SimplyGo Mini Portable Oxygen Concentrator Product code: 1113600
- Reason / hazard
- Certain serial numbers of the SimplyGo Mini standard battery kit contain a defective chip which can lead the battery to either fail to charge or lose voltage output when in use in the device. If the fault occurs while the device is running on battery power, the SimplyGo Mini will shut down and stop providing oxygen to the user. This may happen at a temperature within the specified operating temperature range of the SimplyGo Mini.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- US Nationwide distribution in the states of GA, ID, KY, NJ, NY and the country of Canada.
- Recall initiated
- 2025-10-07
- Code info
- Product Code: 1116816; UDI-DI: 00606959032385; Serial Numbers: 2438V115830, 2438V115960, 2438V116033, 2438V115899, 2438V115832, 2438V115964, 2438V116034, 2438V115837, 2438V115965, 2438V116040, 2438V115840, 2438V115967, 2438V116044, 2438V115842, 2438V115968, 2438V116045, 2438V115844, 2438V115970, 2438V116046, 2438V115878, 2438V115972, 2438V116047, 2438V115890, 2438V115973, 2438V116048, 2438V115894, 2438V115974, 2438V116049, 2438V115895, 2438V115980, 2438V116050, 2438V115896, 2438V115981, 2438V116051, 2438V115897, 2438V115983, 2438V116052, 2438V115919, 2438V115984, 2438V116226, 2438V115937, 2438V115991, 2438V116229, 2438V115939, 2438V115992, 2438V116233, 2438V115942, 2438V115993, 2438V116249, 2438V115948, 2438V115994, 2438V116253, 2438V115957, 2438V115995, 2438V116258, 2438V117762;
- Product quantity
- 56 units (55 US, 1 OUS)
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Murrysville
- State
- PA
- Country
- United States
- Event id
- 97736