Meridian Bioscience Inc: Meridian Bioscience, alethia c. Difficile, Catalog Number 480050, Nucleic Acid Amplification Test Assay
Read the official notice at FDA (device) →
Get alerts for Meridian Bioscience
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Meridian Bioscience Inc
- Product
- Meridian Bioscience, alethia c. Difficile, Catalog Number 480050, Nucleic Acid Amplification Test Assay
- Reason / hazard
- The cap on the reaction buffer tube may be missing the black rubber gasket resulting in the potential for leakage and reduced buffer volume.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- US Distribution to states of: AL, CO, FL, GA, IL, MD, ME, MT, NH, NJ, OH, TN, TX and WA; and OUS (foreign) countries of: Canada, Italy, and Panama.
- Recall initiated
- 2025-10-07
- Code info
- UDI/DI 00840733102172, Lot Numbers: 480050U036, Exp. 2026-08-01; 480050U037, Exp. 2026-09-04
- Product quantity
- 109 in total
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Cincinnati
- State
- OH
- Country
- United States
- Event id
- 97787
Free daily digest + "My products" watchlist