Philips North America: Azurion R3.0. Model Number 722234. Commercial Name: Azurion 7 M20. INTERVENTIONAL FLUOROSCOPIC XRAY SYSTEM
Read the official notice at FDA (device) →
Get alerts for Philips North America
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Philips North America
- Product
- Azurion R3.0. Model Number 722234. Commercial Name: Azurion 7 M20. INTERVENTIONAL FLUOROSCOPIC XRAY SYSTEM
- Reason / hazard
- Between July 2024 and June 2025, Philips Azurion R3.0 systems with a FlexArm stand were shipped without the Source-to-Skin Distance (SSD) Spacer. The SSD spacer is necessary to meet the FDA minimum source-skin distance requirement of 38 cm for standard fluoroscopy applications.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- U.S.
- Recall initiated
- 2025-10-27
- Code info
- Model Number 722234 Commercial Name: Azurion 7 M20 How to identify: Philips Reference C&R 2025-IGT-BST-016
- Product quantity
- 110
- Voluntary mandated
- FDA Mandated
- City
- Cambridge
- State
- MA
- Country
- United States
- Event id
- 98052