B Braun Medical Inc: IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Vol
Read the official notice at FDA (device) →
Get alerts for B Braun Medical
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- B Braun Medical Inc
- Product
- IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET 15D, 2 CARESITES, 153 IN.; Catalog Number: 490542.
- Reason / hazard
- Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.
- Recall initiated
- 2025-10-29
- Code info
- 1. Catalog Number: 490542; Primary UDI-DI: 04046955270187; Unit of Dose UDI-DI: 04046955270170; Expiration Date: Earliest Exp of Comp or 36mths.
- Product quantity
- 7,008 units
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Bethlehem
- State
- PA
- Country
- United States
- Event id
- 97816