Medline Industries, LP: MEDLINE PERFUSION SYRINGE, REF: DYNJ39223C
Read the official notice at FDA (device) →
Get alerts for Medline Industries
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Medline Industries, LP
- Product
- MEDLINE PERFUSION SYRINGE, REF: DYNJ39223C
- Reason / hazard
- Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing BD Luer Tip Caps. BD has issued a recall of their Luer Tip Caps after confirming through internal testing that certain BD Luer Tip Cap trays failed routine biocompatibility testing. Although the tip cap covers a small surface area of the syringe and does not come in direct contact with patients, there is the potential that the failure of a biocompatibility test may have varying consequences ranging from no harm or clinical impact to potential for sensitization, irritation, pyrogenicity, acute systemic toxicity or hemolysis depending on the nature and duration of contact with the syringe.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- US Nationwide distribution in the state of CA.
- Recall initiated
- 2025-10-27
- Code info
- UDI/DI 10193489621532 (each) 40193489621533 (case), Lot Number 25FMJ081
- Product quantity
- 15 units
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Northfield
- State
- IL
- Country
- United States
- Event id
- 97968