Steris Corporation: AMSCO 7053HP Washer/Disinfector Model/Catalog Number: 7053HP Product Description: The AMSCO 7053HP Washer/Disinfector is intended for
Read the official notice at FDA (device) →
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Steris Corporation
- Product
- AMSCO 7053HP Washer/Disinfector Model/Catalog Number: 7053HP Product Description: The AMSCO 7053HP Washer/Disinfector is intended for use in the cleaning and intermediate-level disinfection of soiled reusable simple hard-surfaced rigid surgical instruments (such as forceps and clamps), utensils (such as bedpans, urinals, trays, bowl, basin and kidney dishes), rubber and plastic goods, theatre shoes and other similar and related articles found in healthcare facilities.
- Reason / hazard
- Wire connected to the electrical box may shift out of its intended position, which may result in electrical arcing. Electrical arcing remains internal within the unit (no risk of shock to users).
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- Worldwide - US Nationwide distribution in the states of Arizona, Colorado, Florida, Georgia, Illinois, Kentucky, Louisiana, Michigan, New Hampshire, New York, North Carolina, Pennsylvania, Tennessee, Texas, Utah, Victoria, Virginia, Washington, Wisconsin and the countries of Australia.
- Recall initiated
- 2025-11-19
- Code info
- Lot Code: UDI: 00724995176662, 00724995176679, 00724995176655 Serial numbers of distributed units: 3619725008, 3619725011, 3619725013, 3621625010, 3621625011, 3621625013, 3621625014, 3621625015, 3621925001, 3621925002, 3621925005, 3621925010, 3622425005, 3622725005, 3622725011, 3622725012, 3623125006, 3623425016, 3623825008, 3624025002, 3624025008, 3624525007
- Product quantity
- 22
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Mentor
- State
- OH
- Country
- United States
- Event id
- 97986