Mindray DS USA, Inc. dba Mindray North America: Hardware configuration of the BeneVision Central Monitoring System (CMS), marketed as BeneVision Distributed Monitoring System (DMS). Model
Read the official notice at FDA (device) →
Get alerts for Mindray North America
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Mindray DS USA, Inc. dba Mindray North America
- Product
- Hardware configuration of the BeneVision Central Monitoring System (CMS), marketed as BeneVision Distributed Monitoring System (DMS). Model Number: 115-050935-00 and E115-050935-00. Software Version: Audio Driver version 6.0.9484.1 HP Pro Mini 400 G9 Desktop PC with Realtek HD Audio Driver Officially Released by HP (audio driver version 6.0.9484.1)
- Reason / hazard
- When the Worstation of the BeneVision DMS has a specific hardware configuration, the computer may experience audio playback failure or screen freezing.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- Worldwide distribution - US Nationwide and the country of Canada.
- Recall initiated
- 2025-11-18
- Code info
- Audio Driver Version: 6.0.9484.1. UDI-DI: 06936415972417. HP Pro Mini 400 G9 Desktop PC. Realtek HD Audio Driver Released by HP. 1. WS host package (mini PC). Part Number 115-050935-00. Quantity 902 2. Exch, Benevision Mini PC Workstation. Part Number E115-050935-00. Quantity 4
- Product quantity
- 906 units
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Mahwah
- State
- NJ
- Country
- United States
- Event id
- 97969