AGFA Healthcare Corp.: Brand name: AGFA Digital Radiography X- Ray system DR 800 with MUSICA Dynamic Common Name: DR 800
Read the official notice at FDA (device) →
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- AGFA Healthcare Corp.
- Product
- Brand name: AGFA Digital Radiography X- Ray system DR 800 with MUSICA Dynamic Common Name: DR 800
- Reason / hazard
- It is possible to exceed the maximum allowed Air Kerma rate (AKR) of 88mGy/min for pulsed fluoro exams. This can happen in case the framerate is increased on the NX workstation when ABS is not enabled.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- U.S. Nationwide distribution.
- Recall initiated
- 2025-11-18
- Code info
- N/A
- Product quantity
- 35
- Voluntary mandated
- FDA Mandated
- City
- Greenville
- State
- SC
- Country
- United States
- Event id
- 98101