ICU Medical, Inc.: LifeShield Drug Library Management (DLM), LifeShield Infusion Safety Software Suite v2.2 Product No. 17003-02
Read the official notice at FDA (device) →
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- ICU Medical, Inc.
- Product
- LifeShield Drug Library Management (DLM), LifeShield Infusion Safety Software Suite v2.2 Product No. 17003-02
- Reason / hazard
- Software issue only allows Concentration Limits to be defined to one digit of precision past the decimal point (0.1) instead of three digits (0.001). If the user is unable to use Concentration Limits as intended, they may need to change the limits to something other than what is recommended by the drug manufacturer, which may result in over- or under-delivery.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- US Nationwide distribution in the states of CA, IA, KY, LA, MD, MS, NY, PA, TN, VA.
- Recall initiated
- 2025-11-24
- Code info
- LifeShield Infusion Safety Software Suite v2.2 UDI-DI 00887709127227
- Product quantity
- 12 customers
- Voluntary mandated
- Voluntary: Firm initiated