Roche Diagnostics Operations, Inc.: The Elecsys Anti-TSHR immunoassay is a three-step competitive immunoassay with streptavidin-coated microparticles and electrochemiluminescen
Read the official notice at FDA (device) →
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Roche Diagnostics Operations, Inc.
- Product
- The Elecsys Anti-TSHR immunoassay is a three-step competitive immunoassay with streptavidin-coated microparticles and electrochemiluminescence detection. Patient serum sample is incubated with pretreatment reagent PT1 and the reconstituted PT2 containing a pre-formed immunocomplex of solubilized porcine TSH receptor (pTSHR) and biotinylated anti-pTSHR mouse monoclonal antibody.
- Reason / hazard
- Issues identified: 1) Results vary by Instrument Platform; 2) Calibrator Lot-to-Lot Variability. Issues may lead to delayed or incorrect diagnosis, as well as delay of treatment or incorrect treatment.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- US Nationwide distribution in the states of AZ, CA, FL, GA, IL, IN, ME, MI, MN, NC, NJ, NM, NY, OH, PA, PR, TN, TX, UT, WA, WV.
- Recall initiated
- 2025-11-21
- Code info
- 08496633190 lot 840183 exp. 02/26 08496633190 lot 874011 exp. 10/26 08496609190 lot 840177 exp. 05/26
- Product quantity
- 1300 cassettes
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Indianapolis
- State
- IN
- Country
- United States
- Event id
- 98103