⚠ RecallRadar📷 ScanGuidesOfficial U.S. recall notices, restated plainly

Boston Scientific Corporation: Boston Scientific Coyote OVER-THE-WIRE PTA Balloon Dilation Catheter, 4F, 1) 2.5mm x120mm, 150 cm, UPM/Model No. REF H74939186251210; 2)

Medical DevicesSource: FDA (device) · Date: 2026-01-21

Read the official notice at FDA (device) →

Share on XFacebookRedditEmail

Get alerts for Boston Scientific

Guides: Which agency handles which recall (FDA vs USDA vs CPSC vs NHTSA)

The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.

Company / brand
Boston Scientific Corporation
Product
Boston Scientific Coyote OVER-THE-WIRE PTA Balloon Dilation Catheter, 4F, 1) 2.5mm x120mm, 150 cm, UPM/Model No. REF H74939186251210; 2) 3.5mm x120mm, 150 cm, UPM/Model No. REF H74939186351210; Catheter, Angioplasty, Peripheral, Transluminal
Reason / hazard
Boston Scientific is initiating the removal of certain batches of Coyote" Over-the-Wire (OTW) Percutaneous Transluminal Angioplasty (PTA) Balloon Dilatation Catheter devices due to the potential for inflation/deflation difficulty during use in the peripheral vasculature. An internal investigation found that a manufacturing process variation resulted in catheters that may not properly inflate/deflate. This was limited to Coyote OTW batches manufactured from 12Sep2025 through 13Oct2025. The processing variation has since been addressed. No other material numbers or batches are impacted by this removal.
Classification
Class II
Status
Ongoing
Distribution / sold at
US Nationwide distribution in the states of MA, WV, OH, CA, MN, NY, VA, IL TX, FL, IL.
Recall initiated
2025-12-03
Code info
1) GTIN 08714729796527, Batch Number 37370117, exp. 3-Sept-28; 2) GTIN 08714729796756, Batch Number 37416646, exp. 9-Sept-28
Product quantity
21 units
Voluntary mandated
Voluntary: Firm initiated
City
Maple Grove
State
MN
Country
United States
Event id
98149
Free daily digest + "My products" watchlist

Get the daily digest, and tell us what to watch. Everything is optional except your email.

Free allergen alerts:

Categories:

Privacy: we store only what you enter here, use it only to send recall alerts you asked for, never sell or share it, and you can unsubscribe anytime. Double opt-in: we will email you a confirmation link before sending anything; alerts start only after you confirm. Every email has unsubscribe and manage-watchlist links.