⚠ RecallRadar📷 ScanGuidesOfficial U.S. recall notices, restated plainly

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.: Allura Xper FD10; Model Number: 722003;

Medical DevicesSource: FDA (device) · Date: 2026-01-21

Read the official notice at FDA (device) →

Share on XFacebookRedditEmail

Get alerts for Philips Medical Systems Nederland B.v

Guides: Which agency handles which recall (FDA vs USDA vs CPSC vs NHTSA)

The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.

Company / brand
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Product
Allura Xper FD10; Model Number: 722003;
Reason / hazard
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
Classification
Class II
Status
Ongoing
Distribution / sold at
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Cayman Islands, Chile, China, Colombia, Costa Rica, Cuba, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, France, Germany, Greece, Guatemala, Honduras, India, Italy, Japan, Kenya, Korea, Republic of, Malaysia, Mexico, Nepal, Netherlands, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Qatar, Romania, Russian Federation, Saudi Arabia, Singapore, South Africa, Spain, Sri Lanka, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Venezuela, Yemen.
Recall initiated
2025-12-15
Code info
Model Number: 722003; System Product Name: Allura Xper FD10; UDI-DI: N/A; Serial Numbers: 1232, 967, 746, 1276, 232, 385, 1663, 1189, 81, 700, 167, 1072, 994, 150, 135, 1268, 168, 855, 213, 60, 471, 153, 319, 1159, 162, 180, 142, 1205, 719, 926, 426, 1413, 798, 65, 1165, 1658, 1193, 1778, 2933, 1585, 1380, 1706, 1613, 1685, 1297, 1388, 1522, 1238, 1416, 1448, 536, 1090, 1662, 1083, 1433, 1399, 62, 943, 249, 1227, 163, 1121, 1119, 889, 861, 1649, PP0200000118, 804, 1527, 468, 1593, 1198, 1022, 708, 1054, 865, 744, 1760, 936, 782, 953, 1035, 1058, 179, 1254, 1112, 389, 1560, 1660, 1013, 25, 215, 27;
Product quantity
93 units
Voluntary mandated
Voluntary: Firm initiated
City
Best
Country
Netherlands
Event id
98109
Free daily digest + "My products" watchlist

Get the daily digest, and tell us what to watch. Everything is optional except your email.

Free allergen alerts:

Categories:

Privacy: we store only what you enter here, use it only to send recall alerts you asked for, never sell or share it, and you can unsubscribe anytime. Double opt-in: we will email you a confirmation link before sending anything; alerts start only after you confirm. Every email has unsubscribe and manage-watchlist links.