Heraeus Medical GmbH (Dental Division): Brand Name: PALACOS Product Name: PALACOS MV+G pro 40; PALACOS MV+G pro 80 Model/Catalog Number: 5081290; 5081291 Software Version: N/A
Read the official notice at FDA (device) →
Get alerts for Heraeus Medical GmbH
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Heraeus Medical GmbH (Dental Division)
- Product
- Brand Name: PALACOS Product Name: PALACOS MV+G pro 40; PALACOS MV+G pro 80 Model/Catalog Number: 5081290; 5081291 Software Version: N/A Product Description: Radiopaque, poly(methyl methacrylate)-based (PMMA) bone cement, pre-filled into a mixing and application system, suitable for use with or without vacuum (ready to mix). Component: N/A
- Reason / hazard
- increase of complaints regarding the breakage of the ampoules within their device systems. An extensive investigation revealed that multiple factors have an impact on the ampoule breakage. An unfortunate combination of these influence factors can lead to unbreakable ampoules. If none or not all ampoules break, the product cannot be used on the patient since no or no appropriate cement dough is formed.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- International distribution in the countries of Aland Islands, Australia, Belgium, Denmark, Finland, Germany, Ireland, Italy, Liechtenstein, Netherlands, New Zealand, Norway, Portugal, Spain, Sweden, Switzerland, United Kingdom.
- Recall initiated
- 2025-12-09
- Code info
- Lot Code: UDI-DI: 04260102136243; 04260102136274
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Wehrheim
- Country
- Germany
- Event id
- 98078