Medline Industries, LP: Medline medical procedure convenience kits labeled as: KIT ROBOTICS UROLOGY PROSTATE, REF DYKMBNDL116G
Read the official notice at FDA (device) →
Get alerts for Medline Industries
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Medline Industries, LP
- Product
- Medline medical procedure convenience kits labeled as: KIT ROBOTICS UROLOGY PROSTATE, REF DYKMBNDL116G
- Reason / hazard
- Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- US Nationwide distribution.
- Recall initiated
- 2025-12-16
- Code info
- UDI/DI 10198459156991 (EA) 40198459156992 (CS), Lot Number 25EBM426
- Product quantity
- 12 units
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Northfield
- State
- IL
- Country
- United States
- Event id
- 98197