⚠ RecallRadar📷 ScanGuidesOfficial U.S. recall notices, restated plainly

ICU Medical, Inc.: ICU Medical Plum Solo Precision IV Pump, List Numbers: 400010401 (US), 400011001 (Philippines)

Medical DevicesSource: FDA (device) · Date: 2026-01-28

Read the official notice at FDA (device) →

Share on XFacebookRedditEmail

Get alerts for ICU Medical

Guides: Which agency handles which recall (FDA vs USDA vs CPSC vs NHTSA)

The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.

Company / brand
ICU Medical, Inc.
Product
ICU Medical Plum Solo Precision IV Pump, List Numbers: 400010401 (US), 400011001 (Philippines)
Reason / hazard
Plum Solo and Duo Infusion pumps include a feature, that when selected, automatically flushes the downstream line after a piggyback therapy. The programmed flush volume is delivered from the primary line container at the piggyback therapy rate after the piggyback therapy is complete. In certain cases, as described below, the clinician may receive an Upper Hard Limit (UHL) or Lower Hard Limit (LHL) Violation message when programming a piggyback flush, which prevents the flush from being programmed.
Classification
Class II
Status
Ongoing
Distribution / sold at
Worldwide distribution - US Nationwide and the country of Philippines.
Recall initiated
2025-12-15
Code info
400010401 (US), Software Versions: 1.0.2; 400011001 (Philippines), Software Versions: 1.0.0
Voluntary mandated
Voluntary: Firm initiated
City
Lake Forest
State
IL
Country
United States
Event id
98128
Free daily digest + "My products" watchlist

Get the daily digest, and tell us what to watch. Everything is optional except your email.

Free allergen alerts:

Categories:

Privacy: we store only what you enter here, use it only to send recall alerts you asked for, never sell or share it, and you can unsubscribe anytime. Double opt-in: we will email you a confirmation link before sending anything; alerts start only after you confirm. Every email has unsubscribe and manage-watchlist links.