⚠ RecallRadar📷 ScanGuidesOfficial U.S. recall notices, restated plainly

Canon Medical System, USA, INC.: Alphenix INFX-8000C, interventional fluoroscopic x-ray system

Medical DevicesSource: FDA (device) · Date: 2026-02-04

Read the official notice at FDA (device) →

Share on XFacebookRedditEmail

Get alerts for Canon Medical System

Guides: Which agency handles which recall (FDA vs USDA vs CPSC vs NHTSA)

The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.

Company / brand
Canon Medical System, USA, INC.
Product
Alphenix INFX-8000C, interventional fluoroscopic x-ray system
Reason / hazard
It has been discovered that there is a possibility that the fixing screws of the ceiling movement gear in the ceiling suspended arm that are part of the X-ray interventional systems may become loose. As a result, ceiling lateral movement may no longer be possible, abnormal noise may be generated from the suspension system and a sensor error may be displayed
Classification
Class II
Status
Ongoing
Distribution / sold at
Worldwide distribution - US Nationwide and the country of Dominican Republic.
Recall initiated
2025-12-19
Code info
serial numbers: W1C0632284, W4A1282279, W4A12Z2324, A2A2222381, A2A2392756, A3517001, A2A2352683, A2A2072087, A2A2072077, C1C2422647, W2A09Y2061, W2A10X2168, W4A11Y2144, A2A2232406, A2A2032038, W1D07Z2458, W2A09X2053, B2B1812645, A1C19Z2182, W4B1332362, B2A1622324, W4C1352385, W4A1292284, A1C19X2152, A1C1942058, A1C1972104, B2A1592233, B2A1792598, W2A1012081, W2A1052121, W4C1342374, B2B1872766, A4562105, W2A0972027, A1C2052199, A2A21X2339, W2A1012082, W4C1342369, W1D07Z2449, A1C1942065, A2A2132228, A2A2172285, A2A2117001, W1D06Z2357, A2A2282477, A2A2232405, W1B08X2034, A2A2182313, W4A1232210, A2A2262452, B2A16X2449, B2B1862743, B2B1872751, W1D0697014, W1C05X2229, B2A1542162, A2A21Z2364, A2A2392751, B2A1722527, A2A22Z2566, B4582115, W4C1372400, A2A2362690, A2A2252436, A2A2292510, B2A1582209, B2A1472014, W1D0732376, A2A2252439, W1D0732374, A1C1972105, A1C1972112, B4602126, W1B08Z2063, A2A2152265, A2A2132230, A2A2282491, A2A2262445, A2A22Y2550, A2A22X2528, W4A1212180, W4A1212176, W4A1212177, A1C1992141, W4A1182110, W4A11Z2163, W1D0812466, B2B18Y2816, B2B1832693, A1C1952078, A2A2062070, W2A09Z2077, A1C19Y2174, W1D0722372, A2A22X2518, A3542008, B2A1472020, W4A1242224, W4A1312348, A2A2172280, W4D1422466, W4A1142069, A2A2172287, A2A2342657, A2A2432872, B2A1552176, B2B1812657, B2B1872765, W4A1342373, B2A1682407, W4A11Z2170, W4C1342364, W4C1342365, W4A1172092, A2A2172281, W2A1072143, W4A1292281, A2A2242426, A2A2392761, A3612061, A1A2162240, A2A23Z2814, B2B1872760, A2A2262456, A1C2042197, B2A14Z2085, W4C1352377, W4A1252245, A1C19Z2178, A2A2292499, W4A1242221, W1D0872524, A1C19Z2188, A1C19X2158, A2A2052053, B2B1862731, B2A1752550, B2B1832678, B2B1852708, B2B1852710, B2B1852707, W1C0582207, W1C05Y2242, W1D0722370, W4A1272267, W2A1062127, B2A1742539, B2B1922842, A2A2062066, B2A16Y2469, W4A1122048, B2A1662369, W4A1122039, B5512160, A2A2122205, B2B1822664, B2B1822670, A2A2292498, W4D13Y2442, A2A2122197, A2A2192330, A2A2132216, A2A22X2522, A2A2432885, A2A2422855, W2A1012091, W1C05X2224, B2A1622310, A1C1972111, B2B1892787, A1C1952080, A1B18Y7002.
Product quantity
172 units
Voluntary mandated
Voluntary: Firm initiated
City
Tustin
State
CA
Country
United States
Event id
97970
Free daily digest + "My products" watchlist

Get the daily digest, and tell us what to watch. Everything is optional except your email.

Free allergen alerts:

Categories:

Privacy: we store only what you enter here, use it only to send recall alerts you asked for, never sell or share it, and you can unsubscribe anytime. Double opt-in: we will email you a confirmation link before sending anything; alerts start only after you confirm. Every email has unsubscribe and manage-watchlist links.