Medline Industries, LP: Medline medical procedure convenience kits labeled as: 1) INPATIENT CTRL LN DRG CHG, Kit SKU DT22880; 2) ADULT CVC DRSG CHG W/ADHESIVE,
Read the official notice at FDA (device) →
Get alerts for Medline Industries
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Medline Industries, LP
- Product
- Medline medical procedure convenience kits labeled as: 1) INPATIENT CTRL LN DRG CHG, Kit SKU DT22880; 2) ADULT CVC DRSG CHG W/ADHESIVE, Kit SKU DYNDC3283A.
- Reason / hazard
- Medline kits were manufactured and distributed containing MASTISOL liquid adhesive which was recalled due to complaints that the butyrate tubing cracked during actuation of the vials of MASTISOL Liquid Adhesive.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- US Nationwide distribution.
- Recall initiated
- 2025-12-12
- Code info
- Medline Kit SKU DT22880: UDI/DI 10653160995470 (EA) 40653160995471 (CS), Lot Number 25KMJ441; Medline Kit SKU DYNDC3283A: UDI/DI 10198459313240 (EA) 40198459313241 (CS), Lot Number 25KMH515.
- Product quantity
- 1,928 total
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Northfield
- State
- IL
- Country
- United States
- Event id
- 98226