Olympus Corporation of the Americas: Brand Name: Single Use Electrosurgical Knife Product Name: Olympus Single Use Electrosurgical Knife KD-640L Model/Catalog Number: KD640-L
Read the official notice at FDA (device) →
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Olympus Corporation of the Americas
- Product
- Brand Name: Single Use Electrosurgical Knife Product Name: Olympus Single Use Electrosurgical Knife KD-640L Model/Catalog Number: KD640-L Product Description: An electrosurgical device consisting of a handpiece with mechanical and/or electrical controls, and a monopolar electrode, intended to deliver a current from a system generator directly to tissues for cutting/coagulation/ablation during endoscopic (including laparoscopic, arthroscopic) surgery.
- Reason / hazard
- A deterioration of the cutting knife, including overheating and burning, can contribute to tip breakage during use.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- US Nationwide distribution.
- Recall initiated
- 2024-02-17
- Code info
- Model/Catalog Number: KD640-L UDI: 04953170208423 Lot Number(s): 2ZK, 31K, 32K, 33K, 34K, 35K, 36K, 37K, 38K, 39K, 3XK, 3YK, 3ZK, 41K, 42K, 43K, 44K, 45K, 46K, 47K, 48K, 49K, 4XK, 4YK, 4ZK, 51K, 52K, 53K, 54K, 55K, 58K, 59K
- Product quantity
- units
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Center Valley
- State
- PA
- Country
- United States
- Event id
- 98007