Beckman Coulter, Inc.: CellMek SPS Sample Preparation System, REF: C44603, with software
Read the official notice at FDA (device) →
Get alerts for Beckman Coulter
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Beckman Coulter, Inc.
- Product
- CellMek SPS Sample Preparation System, REF: C44603, with software
- Reason / hazard
- Sample preparation system designed to automate staining, lysing, cell washing has a software error that causes samples to be dispensed close to or touching tube bottoms, which could dislodge fluid line, result in sample loss, spillage, make instrument inoperable and will lead to unintentional removal of 60-100 microliters from final sample, may lead to incorrect diagnosis or patient management.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- Worldwide - US Nationwide distribution in the states of MO, WA, MA, IL, AL, OR, IL and the countries of AB, BE, ES, FR, NL, ON, DE, AT, VIC, AU, GB, SK, MX, CH.
- Recall initiated
- 2025-12-23
- Code info
- UDI-DI: 15099590750312. Software Version 2.3.90.0
- Product quantity
- 50
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Miami
- State
- FL
- Country
- United States
- Event id
- 98207