Vortex Surgical Inc.: Vortex Surgical Lindsell Sutured IOL Marker, VS0390 - Scleral Marker

Medical DevicesSource: FDA (device) · Date: 2026-02-11

Read the official notice at FDA (device) →

Share on XFacebookRedditEmail

Get alerts for Vortex Surgical

Guides: Which agency handles which recall (FDA vs USDA vs CPSC vs NHTSA)

The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.

Company / brand
Vortex Surgical Inc.
Product
Vortex Surgical Lindsell Sutured IOL Marker, VS0390 - Scleral Marker
Reason / hazard
There may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots. A compromised sterile barrier has potential of bioburden contamination which could lead to infection.
Classification
Class II
Status
Ongoing
Distribution / sold at
Worldwide Distribution: US (nationwide) to states of: AL, AR, AZ, CA, CT, FL, HI, IA, IL, IN, KY, MA, MI, MN, MS, NC, NE, NJ, NM, NY, OH, OR, PA, RI, SC, SD, TN, TX, WA, WI, VA; and OUS (foreign) to countries of: Australia, Chile, China, France, Germany, India, Ireland, Italy, Japan, Korea, Mexico, Netherlands, Philippines, Poland, Puerto Rico, Spain, Switzerland, Taiwan, United Kingdom, Vietnam.
Recall initiated
2025-12-16
Code info
Pouch UDI 810123483655 Box UDI 810123483662 Lot 2408035
Product quantity
14,789 (8651 US; 6138 OUS)
Voluntary mandated
Voluntary: Firm initiated
City
Saint Charles
State
MO
Country
United States
Event id
98255
Free daily digest + "My products" watchlist

Get the daily digest, and tell us what to watch. Everything is optional except your email.

Free allergen alerts:

Categories:

Privacy: we store only what you enter here, use it only to send recall alerts you asked for, never sell or share it, and you can unsubscribe anytime. Double opt-in: we will email you a confirmation link before sending anything; alerts start only after you confirm. Every email has unsubscribe and manage-watchlist links.