Medline Industries, LP: Intubation Tray (Adult) containing Suction Catheter Kits SKU DYNJAA245B
Read the official notice at FDA (device) →
Get alerts for Medline Industries
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Medline Industries, LP
- Product
- Intubation Tray (Adult) containing Suction Catheter Kits SKU DYNJAA245B
- Reason / hazard
- A smaller open suction catheter, the 8 Fr Suction Catheter, may have been packed within the kit instead of the required 14 Fr Suction Catheter. Use of the smaller size catheter may result in hypoxia and/or respiratory failure related to partial or full airway obstruction. Airway obstruction may require immediate emergent medical intervention to prevent respiratory failure or death. Other potential outcomes include prolonged procedure, airway irritation, or prolonged dyspnea.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- Distribution US nationwide and Canada.
- Recall initiated
- 2025-12-24
- Code info
- Kit lots 25IBC172 25HBJ413 UDI (ea) 10198459480317 UDI (case) 40198459480318
- Product quantity
- 130
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Northfield
- State
- IL
- Country
- United States
- Event id
- 98285