Medline Industries, LP: Reprocessed Abbott Inquiry Steerable Diagnostic Catheters: Abbott Inquiry Steerable Diagnostic Catheter 10 Elec. 2-5-2mm 81105RH 8117
Read the official notice at FDA (device) →
Get alerts for Medline Industries
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Medline Industries, LP
- Product
- Reprocessed Abbott Inquiry Steerable Diagnostic Catheters: Abbott Inquiry Steerable Diagnostic Catheter 10 Elec. 2-5-2mm 81105RH 81172RH; Abbott Inquiry Steerable Diagnostic Catheter 20 Elec. 2-10-2mm 81202RH; Abbott Inquiry Steerable Diagnostic Catheter 4 Elec. 5mm 81483RH; Inquiry Steerable Diagnostic Catheter 10 Elec. 2-5-2mm 81102RH 81104RH 81174RH 81531RH 81532RH; Inquiry Steerable Diagnostic Catheter 10 Elec. 5mm 81107RH; Inquiry Steerable Diagnostic Catheter 4 Elec. 2-5-2mm 81402RH 81404RH 81405RH 81474RH; Inquiry Steerable Diagnostic Catheter 4 Elec. 5mm 81473RH
- Reason / hazard
- These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus reaction which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Expansion of recall Z-2610/2614-2025
- Classification
- Class I
- Status
- Ongoing
- Distribution / sold at
- Distribution US nationwide.
- Recall initiated
- 2025-12-22
- Code info
- 81105RH UDI-DI 10197344133253 Lot EP250618; 81172RH UDI-DI 10197344133284 Lot EP250618; 81202RH UDI-DI 10197344133307 Lot EP250618; 81483RH UDI-DI 10197344133444 Lot EP250618; 81102RH UDI-DI (case) 20197344133236¿ (ea) 10197344133239 Lot EP250618; 81104RH UDI-DI 10197344133246 Lot EP250618; 81174RH UDI-DI 10197344133291 Lot EP250618; 81531RH UDI-DI 10197344133512 Lot EP250618; 81532RH UDI-DI 10197344133529 Lot EP250618; 81107RH UDI-DI 10197344133260 Lot EP250618; 81402RH UDI-DI 10197344133352 Lot EP250618; 81404RH UDI-DI 10197344133376 Lot EP250618; 81405RH UDI-DI 10197344133383 Lot EP250618; 81474RH UDI-DI 10197344133437 Lot EP250618; 81473RH UDI-DI 10197344133420 Lot EP250618
- Product quantity
- 311
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Northfield
- State
- IL
- Country
- United States
- Event id
- 98277