Laerdal Medical Corporation: Laerdal Compact Suction Unit 4 RTCA version (LCSU 4 RTCA) 1) LCSU 4, 800 ml, RTCA; Catalog Number: 880052; 2) LCSU 4, 300 ml, RTCA; Ca
Read the official notice at FDA (device) →
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Laerdal Medical Corporation
- Product
- Laerdal Compact Suction Unit 4 RTCA version (LCSU 4 RTCA) 1) LCSU 4, 800 ml, RTCA; Catalog Number: 880052; 2) LCSU 4, 300 ml, RTCA; Catalog Number: 880062; 3) LCSU 4, RTCA; Catalog Number: 881152;
- Reason / hazard
- units manufactured in a limited time period may emit electromagnetic noise beyond the acceptable limits for RTCA application specified in RTCA DO-160G Chapter 21, Equipment Category M.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- Worldwide distribution - US Nationwide and the countries of Denmark, Sweden, Norway, Canada, Finland, Australia, Spain, France, United Kingdom, Italy, Korea, Germany, Netherlands, Switzerland, New Zealand, Singapore, Malaysia, Austria.
- Recall initiated
- 2026-01-13
- Code info
- 1) LCSU 4, 800 ml, RTCA; Catalog Number: 880052; UDI-DI: 07045432088519; 2) LCSU 4, 300 ml, RTCA; Catalog Number: 880062; UDI-DI: 07045432088533; 3) LCSU 4, RTCA; Catalog Number: 881152; UDI-DI: 07045432067132;
- Product quantity
- 1202 units (537 US, 665 OUS) (2/23/2026: 574 US, 1239 units)
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Wappingers Falls
- State
- NY
- Country
- United States
- Event id
- 98039