⚠ RecallRadar📷 ScanGuidesOfficial U.S. recall notices, restated plainly

Medtronic Perfusion Systems: Octopus Nuvo Tissue Stabilizer, disposable tissue stabilization system, Product Number TSMICS1

Medical DevicesSource: FDA (device) · Date: 2026-03-04

Read the official notice at FDA (device) →

Share on XFacebookRedditEmail

Get alerts for Medtronic Perfusion Systems

Guides: Which agency handles which recall (FDA vs USDA vs CPSC vs NHTSA)

The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.

Company / brand
Medtronic Perfusion Systems
Product
Octopus Nuvo Tissue Stabilizer, disposable tissue stabilization system, Product Number TSMICS1
Reason / hazard
During the manufacturing, Medtronic personnel identified an assembly issue where the tubing was incorrectly positioned. The short tube was attached to the bottom of the canister, and the long tube was attached to the top, contrary to the intended design.
Classification
Class II
Status
Ongoing
Distribution / sold at
US Nationwide. Global Distribution.
Recall initiated
2026-01-13
Code info
UDI-DI: 00763000543693; Lot Number: 0232837110, Additional lots as of 5/22/26: 0231547054 0231591827 0231758880 0231758898 0231758932 0231792414 0231792421 0231912903 0232670901 0232709219 0232709254 0232744928 0232784925 0232784989 0232805458
Product quantity
939 units
Voluntary mandated
Voluntary: Firm initiated
City
Brooklyn Park
State
MN
Country
United States
Event id
98354
Free daily digest + "My products" watchlist

Get the daily digest, and tell us what to watch. Everything is optional except your email.

Free allergen alerts:

Categories:

Privacy: we store only what you enter here, use it only to send recall alerts you asked for, never sell or share it, and you can unsubscribe anytime. Double opt-in: we will email you a confirmation link before sending anything; alerts start only after you confirm. Every email has unsubscribe and manage-watchlist links.