Medtronic Perfusion Systems: Octopus Nuvo Tissue Stabilizer, disposable tissue stabilization system, Product Number TSMICS1
Read the official notice at FDA (device) →
Get alerts for Medtronic Perfusion Systems
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Medtronic Perfusion Systems
- Product
- Octopus Nuvo Tissue Stabilizer, disposable tissue stabilization system, Product Number TSMICS1
- Reason / hazard
- During the manufacturing, Medtronic personnel identified an assembly issue where the tubing was incorrectly positioned. The short tube was attached to the bottom of the canister, and the long tube was attached to the top, contrary to the intended design.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- US Nationwide. Global Distribution.
- Recall initiated
- 2026-01-13
- Code info
- UDI-DI: 00763000543693; Lot Number: 0232837110, Additional lots as of 5/22/26: 0231547054 0231591827 0231758880 0231758898 0231758932 0231792414 0231792421 0231912903 0232670901 0232709219 0232709254 0232744928 0232784925 0232784989 0232805458
- Product quantity
- 939 units
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Brooklyn Park
- State
- MN
- Country
- United States
- Event id
- 98354
Free daily digest + "My products" watchlist