Olympus Corporation of the Americas: Brand Name: ShockPulse-SE Lithotripsy System SPL-S Product Name: ShockPulse-SE Lithotripsy System - Single Use Probes Model/Catalog Number
Read the official notice at FDA (device) →
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Olympus Corporation of the Americas
- Product
- Brand Name: ShockPulse-SE Lithotripsy System SPL-S Product Name: ShockPulse-SE Lithotripsy System - Single Use Probes Model/Catalog Number: SPL-S Product Description: The ShockPulse Lithotripsy System is an electromechanical device capable of fragmenting calculi and aspirating stone debris. The ShockPulse transducer is a reusable device that requires cleaning and steam sterilization prior to use in order to generate both low-frequency mechanical impacts and high-frequency ultrasonic energy that quickly disintegrates stones.
- Reason / hazard
- Ongoing investigations identified additional instances of the device Generator remaining in a blinking phase waiting to recognize the transducer. Damage to the transducer plug and/or generator receptacle may cause these issues.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- Worldwide distribution - US Nationwide and the countries of Canada, Mexico, Bolivia, Chile, India, China, EU, Singapore, Australia, Korea, and Japan.
- Recall initiated
- 2025-11-14
- Code info
- Model/Catalog Number: SPL-S; UDI: 00821925044197; Serial Numbers: All;
- Product quantity
- 1082 units
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Center Valley
- State
- PA
- Country
- United States
- Event id
- 98359