Instrumentation Laboratory: GEM Premier 5000; Model No. 00055430010.

Medical DevicesSource: FDA (device) · Date: 2026-03-18

Read the official notice at FDA (device) →

Share on XFacebookRedditEmail

Get alerts for Instrumentation Laboratory

Guides: Which agency handles which recall (FDA vs USDA vs CPSC vs NHTSA)

The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.

Company / brand
Instrumentation Laboratory
Product
GEM Premier 5000; Model No. 00055430010.
Reason / hazard
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.
Classification
Class II
Status
Ongoing
Distribution / sold at
Worldwide distribution - US Nationwide and the countries of Argentina, Austria, Australia, Belgium, Bulgaria, Brazil, Belarus, Switzerland, Canada, Chile, China, Colombia, Czech Republic, Germany, Denmark, Dominican Republic, Ecuador, Estonia, Spain, Finland, France, United Kingdom, Georgia, Greece, Guatemala, Hong Kong, Croatia, Hungary, Indonesia, Ireland, Israel, India, Italy, Jordan, Japan, South Korea, Kazakhstan, Lebanon, Lithuania, Luxembourg, Latvia, Macau, Mexico, Malaysia, Netherlands, New Zealand, Oman, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Serbia, Saudi Arabia, Sweden, Singapore, Slovenia, San Marino, South Africa, Thailand, Tunisia, Turkey, Uruguay, Uzbekistan & Vietnam.
Recall initiated
2026-01-20
Code info
Part number: 00055430010; UDI: 08426950807759; Lots No.: 251021K, 251027AA, 251110AH, 251114K, 251114O, 251114P, 251114Y, 251117AA, 251117AC, 251117AG, 251117AM, 251117AN, 251117P, 251117T, 251117V, 251117Y, 251118A, 251118B, 251118R, 251118X, 251119A, 251120R, 251120U, 251121Q, 251124O, 251124Q, 251125I, 251126S, 251126T, 251126U, 251201W, 251201X, 251202G, 251202H, 251203L, 251204R, 251205L, 251208N, 251211W, 251215D, 251215Z, 251216O, 251216S, 251217L, 251217V, 251218AB, 251218B, 251218X, 251218Y, 251219A, 251219O, 251222AE, 251222E, 251222F, 251222K, 251222M, 251222P, 251222Z, 251223A, 251223B, 251223C, 251223I, 251223M, 251223O, 251229D, 251229E, 251229I, 251229S, 251231B, 251231G.
Product quantity
1,607 units
Voluntary mandated
Voluntary: Firm initiated
City
Bedford
State
MA
Country
United States
Event id
98349
Free daily digest + "My products" watchlist

Get the daily digest, and tell us what to watch. Everything is optional except your email.

Free allergen alerts:

Categories:

Privacy: we store only what you enter here, use it only to send recall alerts you asked for, never sell or share it, and you can unsubscribe anytime. Double opt-in: we will email you a confirmation link before sending anything; alerts start only after you confirm. Every email has unsubscribe and manage-watchlist links.