I.T.S. GmbH: I.T.S. Fibula Plate PROlock with Angular Stability with the below descriptions and corresponding article numbers. 1. Distal Tubular Plat
Read the official notice at FDA (device) →
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- I.T.S. GmbH
- Product
- I.T.S. Fibula Plate PROlock with Angular Stability with the below descriptions and corresponding article numbers. 1. Distal Tubular Plate; Article Numbers: 21602-3, 21602-4, 21602-5, 21602-6, 21602-7, 21602-8, 21603-3, 21603-4, 21603-5, 21603-6, 21603-7, 21603-8. 2. Fibula Plate 3.5mm; Article Numbers: 21225-4, 21225-6, 21225-8, 21225-10, 21225-12, 21226-4, 21226-6, 21226-8, 21226-10, 21226-12. 3. Universal Tubular Plate; Article Numbers: 21601-5, 21601-6, 21601-7, 21601-8, 21601-10, 21601-12, 21601-14, 21601-16.
- Reason / hazard
- Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.
- Recall initiated
- 2026-01-23
- Code info
- 1. Article Number (Lot Numbers): 21602-3 (643/072330, 643/092310), 21602-4 (643/072301, 643/072308), 21602-5 (643/022417, 643/0650, 643/072309), 21602-6 (643/042305, 643/072333), 21602-7 (643/072332, 643/082305), 21602-8 (43/0405, 643/012305), 21603-3 (643/072331, 643/082306), 21603-4 (643/102214, 643/102206, 643/072302, 643/022313, 643/032310), 21603-5 (643/072335, 643/122308), 21603-6 (643/022309, 643/072305, 643/072311, 643/122309), 21603-7 (643/072337, 643/082313), 21603-8 (643/072312, 643/122204). 2. Article Number (Lot Numbers): 21225-4 (768/20894, 26/F65041, 26/F66149, 679/092305, 679/022430, 26/F705141, 26/F72362, 26/F73634), 21225-6 (768/20895, 768/20039, 26/F65042, 26/F66150, 26/F66499, 679/092321, 26/F705161, 679/062459, 26/F70516, 26/F73635), 21225-8 (768/20896, 26/F65635, 26/F66500, 26/F70515), 21225-10 (26/F66497), 21225-12 (26/F66498, 679/012438), 21226-4 (26/F65043, 26/F66152, 26/F66501, 26/F70521, 26/F705211, 679/012412, 679/062306, 679/092306, 768/20897, 768/208978), 21226-6 (26/F35433, 26/F65044, 26/F66502, 26/F70522, 26/F705221, 679/062307, 679/092307, 768/20898), 21226-8 (26/F66503, 679/012413, 768/20045, 768/20899), 21226-10 (26/F66151), 21226-12 (26/F66148, 679/012411). 3. Article Number (Lot Numbers): 21601-5 (643/012302, 643/072323), 21601-6 (643/072326, 770/201904), 21601-7 (643/012303, 643/022308, 643/0647, 643/072315, 643/072327, 643/082303), 21601-8 (643/072328, 643/072329), 21601-10 (643/022312, 643/082301, 643/082302), 21601-12 (26/F70342), 21601-14 (26/F70343), 21601-16 (26/F70344).
- Product quantity
- 869 units
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Lasnitzhohe
- Country
- Austria
- Event id
- 98341