Olympus Corporation of the Americas: Brand Name: Olympus PKS Cutting Forceps w/Cord, 5mm 24cm Product Name: PKS Cutting Forceps Model/Catalog Number: 920000PK Product Descrip
Read the official notice at FDA (device) →
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Olympus Corporation of the Americas
- Product
- Brand Name: Olympus PKS Cutting Forceps w/Cord, 5mm 24cm Product Name: PKS Cutting Forceps Model/Catalog Number: 920000PK Product Description: The PKS Cutting Forceps are intended to be passed through a 5mm cannula. Coagulation is achieved using electrosurgical energy under visualization. The device is intended to be used with the PKS electrical generator.
- Reason / hazard
- Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- Worldwide - US Nationwide and the countries of Canada, Mexico, EU, Australia, China, Singapore, Korea, Japan.
- Recall initiated
- 2026-02-25
- Code info
- Model: 920000PK; UDI: 00821925038080; Lot Numbers: All unexpired;
- Product quantity
- 7 units
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Center Valley
- State
- PA
- Country
- United States
- Event id
- 98454