Siemens Medical Solutions USA, Inc: ARTIS pheno and ARTIS Icono biplane, floor and ceiling. Model Numbers 10849000, 11327600, 11328100, 11327700.

Medical DevicesSource: FDA (device) · Date: 2026-04-15

Read the official notice at FDA (device) →

Share on XFacebookRedditEmail

Get alerts for Siemens Medical Solutions

Guides: Which agency handles which recall (FDA vs USDA vs CPSC vs NHTSA)

The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.

Company / brand
Siemens Medical Solutions USA, Inc
Product
ARTIS pheno and ARTIS Icono biplane, floor and ceiling. Model Numbers 10849000, 11327600, 11328100, 11327700.
Reason / hazard
During patient examinations, the system may sporadically display a major increase of the x-ray dose applied to the patient. Despite the reported dose showing significantly higher, the actual applied x-ray dose levels are in accordance with the system settings and imaging conditions, as well as the applicable regulatory requirements.
Classification
Class II
Status
Ongoing
Distribution / sold at
Worldwide - US Nationwide distribution.
Recall initiated
2026-03-10
Code info
ARTIS pheno UDI: 4056869046877. ARTIS icono biplane UDI: 4056869063317. ARTIS icono ceiling UDI: 4056869295923. ARTIS icono floor UDI: 4056869149325.
Product quantity
U.S. 758, OUS 2077
Voluntary mandated
FDA Mandated
City
Malvern
State
PA
Country
United States
Event id
98675
Free daily digest + "My products" watchlist

Get the daily digest, and tell us what to watch. Everything is optional except your email.

Free allergen alerts:

Categories:

Privacy: we store only what you enter here, use it only to send recall alerts you asked for, never sell or share it, and you can unsubscribe anytime. Double opt-in: we will email you a confirmation link before sending anything; alerts start only after you confirm. Every email has unsubscribe and manage-watchlist links.