Medline Industries, LP: THORACIC ROBOTS, DYNJ908777B
Read the official notice at FDA (device) →
Get alerts for Medline Industries
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Medline Industries, LP
- Product
- THORACIC ROBOTS, DYNJ908777B
- Reason / hazard
- Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- US Nationwide distribution in the states of CA, FL, IL, IN, KY, MA, MD, MT, NY, OH, PA, TN, TX.
- Recall initiated
- 2026-02-16
- Code info
- UDI-DI 10198459269196 (EA) 40198459269197 (CS) LOTS 25EBD732 25GBG733 25IBQ613 25JBD457 25KBT352 26ABA235 26ABP735
- Product quantity
- 14,379 kits total
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Northfield
- State
- IL
- Country
- United States
- Event id
- 98530