Civco Medical Instruments Co. Inc.: eTRAX Needle Sensor - 18G (for Aurora Trackers), Part Number 667-159
Read the official notice at FDA (device) →
Get alerts for Civco Medical Instruments
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Civco Medical Instruments Co. Inc.
- Product
- eTRAX Needle Sensor - 18G (for Aurora Trackers), Part Number 667-159
- Reason / hazard
- There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- US States: OH, PA, WA. China.
- Recall initiated
- 2026-03-02
- Code info
- UDI-DI: 00841436111102; Lot Number: A257459, A252198, A268105, A276632
- Product quantity
- 21 units
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Kalona
- State
- IA
- Country
- United States
- Event id
- 98513
Free daily digest + "My products" watchlist