Medtronic MiniMed, Inc.: MiniMed 780G Insulin Pump (MMT-1884, MMT-1885, MMT-1886, MMT-1894, MMT-1895, MMT-1896)

Medical DevicesSource: FDA (device) · Date: 2026-04-08

Read the official notice at FDA (device) →

Share on XFacebookRedditEmail

Get alerts for Medtronic MiniMed

Guides: Which agency handles which recall (FDA vs USDA vs CPSC vs NHTSA)

The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.

Company / brand
Medtronic MiniMed, Inc.
Product
MiniMed 780G Insulin Pump (MMT-1884, MMT-1885, MMT-1886, MMT-1894, MMT-1895, MMT-1896)
Reason / hazard
All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
Classification
Class II
Status
Ongoing
Distribution / sold at
Worldwide - US Nationwide distribution including in the states of OR, CA, AZ, NY,TX,FL, OH, MO, TN,NJ, NC, SC, IL, VA, WA, DC, GA, PA, MS,IA, MI, NH, OK, SD, MT, WI, WV, MA, LA, MN, RI, AR, ME, AL, IN, UT, KY, CO, ND, MD, CT, NM, KS, ID, NV, DE, NE, VT, AK, HI, WY, PR, VI and the countries of Taiwan, New Zealand, Australia, Bangladesh, Brunei Darussalam, Cambodia, Indonesia, Malaysia, Maldives, Philippines, Thailand, Vietnam, Singapore, Pakistan, Hong Kong, China, Argentina, Aruba, Barbados, Bermuda, Bolivia, Plurinational State Of, Brazil, Chile, Colombia, Costa Rica, Dominican Republic, Ecuador, El Salvador, Guatemala, Honduras, Mexico, Panama, Paraguay, Peru, Puerto Rico, Trinidad And Tobago, Uruguay, Venezuela, Canada, South Korea, India, Albania, Algeria, Andorra, Armenia, Austria, Azerbaijan, Bahrain, Belarus, Belgium, Bosnia And Herzegovina, Botswana, Bulgaria, Canary Islands, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Faroe Islands, Finland, France, French Guyana, French Polynes., Georgia, Germany, Gibraltar, Greece, Guadeloupe, Hungary, Iceland, Iran, Iraq, Ireland, Israel, Italy, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Kyrgyzstan, Latvia, Lebanon, Libya, Liechtenstein, Lithuania, Luxembourg, Macedonia, Malawi, Malta, Martinique, Mauritius, Mayotte, Moldavia, Montenegro, Morocco, Namibia, Netherlands, New Caledonia, Nigeria, Northern Ireland, Norway, Oman, Poland, Portugal, Qatar, Reunion, Romania, Russia, San Marino, Saudi Arabia, Serbia, Sierra Leone, Slovakia, Slovenia, South Africa, Spain, St.Pier,Miquel., Sudan, Sweden, Switzerland, Syrian Arab Republic, Tanzania, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Yemen, Zambia, Zimbabwe.
Recall initiated
2026-02-13
Code info
Pump: UDI-DI(GTIN): User Guide Name(Part-Version): MiniMed 780G Insulin Pump: 763000411541, 763000441449, 763000523305, 763000521530, 763000521516, 763000533861, 763000544317, 763000734459, 763000758530, 763000758547, 763000578688, 763000734435, 763000504762, 763000734466, 763000924102, 763000924089, 763000454296, 763000854829, 763000411411, 763000411503, 763000384272, 763000411466, 763000411428, 763000411572, 763000411589, 763000411558, 763000384265, 763000411473, 763000411374, 763000411435, 763000411404, 763000384289, 763000411459, 763000411510, 763000411626, 763000411565, 763000411527, 763000411619, 763000411596, 763000411497, 763000411398, 763000411534, 763000411381, 763000384258, 763000411480, 763000411602, 763000411640, 763000411442, 763000441432, 763000442194, 763000521523, 763000442187, 763000521509, 763000521547, 763000578671, 763000596194, 763000639228, 763000639235, 763000639266, 763000974381, 763000974374, 763000921750, 763000921767, 763000921590, 763000921637, 763000933593, 763000891886, 763000921729, 763000921866, 763000921781, 763000921811, 763000921576, 763000921583, 763000921620, 763000921859, 763000891909, 763000921743, 763000944162, 763000921712, 763000921668, 763000933579, 763000933555, 763000933586, 763000933609, 763000921613, 763000933715, 763000933548, 763000944209, 763000891893, 763000933623, 763000933722, 763000933814, 763000933708, 763000933777, 763000933685, 763000933531, 763000933678, 763000933821, 763000933746, 763000933562, 763000944193, 763000533878, 763000734442, 763000854942, 763000914127: Medtronic MiniMed 780G with Instinct sensor System User Guide(M018856C024 2, M018857C035 2); Medtronic MiniMed 780G with Simplera Sync sensor System User Guide(M018857C034 2); Medtronic MiniMed 780G with Guardian 4 Sensor System User Guide(M028790C004 2, M028792C004 2); Medtronic MiniMed 780G with Guardian 4 Sensor System User Guide(M028792C015 1, M028792C001 3, M028792C002 1, M028792C006 1, M028792C003 1, M028792C005 1); Medtronic MiniMed 780G with Instinct sensor System User Guide(M018857C008 1, M018857C011 1, M018857C029 1, M018856C022 1, M018857C013 1, M018856C007 1, M018857C001 4, M018856C001 4, M018856C010 1, M018857C012 1, M018856C020 1, M018857C009 1, M018857C032 1, M018857C022 1, M018856C021 1, M018856C013 1, M018857C028 1, M018857C033 1, M018856C017 1, M018857C016 1, M018856C018 1); Medtronic MiniMed 780G with Simplera Sync sensor System User Guide(M018856C023 2, M018857C034 2); MiniMed 780G System Technical Guide(M075489C003 1, M075489C002 1, M075489C001 2); MiniMed 780G System User Guide(M072481C001 2); MiniMed 780G with Guardian 4 Sensor System User Guide mg(, M073070C003 1, M073070C001 2, M073070C002 1); MiniMed 780G with Guardian 4 Sensor System User Guide mm(M076019C002 1, M076019C001 1); MiniMed 780G with Guardian 4 Sensor System User Guide(M028790C003 1, M028790C002 1); MiniMed 780G system Changing the Medtronic Extended Infusion Set with Extended Reservoir(M048911C001 2, M048911C003 1, M048911C002 1); MiniMed 780G system Changing the MiniMed Mio Advance Infusion Set and Reservoir(M047607C001 1); MiniMed 780G system Changing the MiniMed Silhouette Infusion Set and Reservoir(M047610C001 1, M047610C003 1, M047610C002 1); MiniMed 780G system Changing the MiniMed Sure-T Infusion Set and Reservoir(M047609C001 1, M047609C003 1, M047607C003 1, M047607C002 1, M047609C002 1); MiniMed 780G System User Guide(M024832C003 2, M024829C004 2, M024832C004 2, M024829C005 2); MiniMed 780G System User Guide mg/dl(M014786C005 2, M014784C005 2, M014786C007 1, M014784C007 1, M014784C004 1, M014786C004 1, M014786C008 1, M014784C008 1, M014784C002 1, M014786C006 1, M014784C006 1, M014784C016 1, M014786C016 1, M014784C012 1, M014786C012 1, M014786C002 1, M014784C010 1, M014786C010 1, M014784C003 1, M014786C003 1, M014784C011 1, M014786C011 1, M014786C013 1, M014784C013 1, M014784C015 1, M014786C015 1, M014786C019 1, M014784C019 1); MiniMed 780G System User Guide mmol/L(M014787C002 1, M014785C002 1, M014785C007 2, M014787C007 1, M014787C005 1, M014785C005 1, M014787C006 1, M014785C006 1, M014785C023 2, M014787C023 2, M014787C010 1, M014785C010 1, M014787C011 1, M014785C011 1, M014787C021 1, M014785C021 1, M014787C008 1, M014785C008 1, M014785C009 1, M014787C009 1, M014787C022 1, M014785C022 1, M014785C027 1, M014787C027 1, M014787C004 2, M014785C004 2, M014785C003 1, M014787C003 1, M014785C030 1, M014787C030 1, M014785C013 2, M014787C013 2, M014785C019 1, M014787C019 1, M014785C020 1, M014787C020 1, M014787C018 1, M014785C018 1, M014787C012 1, M014785C012 1, M014787C016 1, M014785C016 1, M014787C024 1, M014785C024 1, M014785C014 2, M014787C014 1, M014787C028 1, M014785C028 1); MiniMed 780G with Guardian 4 Sensor System User Guide(M060264C012 1, M060263C011 1, M060264C018 1, M060263C015 1, M060264C017 1, M060263C014 1, M060264C006 1, M060264C004 1, M060263C001 3, M060263C016 1, M060264C007 1, M060264C008 1, M060264C009 1, M060264C010 1, M060264C011 1, M060264C001 3, M060264C013 1, M060264C014 1, M060264C023 1, M028790C005 1, M060263C020 1, M060264C015 1, M060264C016 1, M060264C020 1, M060263C021 1, M028790C006 1, M060263C022 1, M056377C001 3, M056377C003 1, M056377C002 1, M060263C023 1, M060264C021 1, M060264C022 1, M060263C024 1); MiniMed 780G with Guardian 4 Sensor System User Guide mg/dl(M028790C001 3, M028790C011 1, M028790C007 1, M028790C008 1, M028790C010 1, M028790C019 1); MiniMed 780G with Guardian 4 Sensor System User Guide mmol/L(M028792C012 1, M028792C013 1, M028792C007 1, M028792C011 1, M028792C025 1, M028792C014 1, M028792C001 3, M028792C033 1, M028792C010 1); MiniMed 780G with Instinct sensor System User Guide(M018857C004 1); MiniMed 780G with Simplera Sync sensor System User Guide(M060262C008 1, M060261C002 1, M018857C003 1, M018856C003 1, M060262C015 1, M018856C005 1, M060261C004 1, M060261C003 1, M018856C006 1, M060262C014 1, M060262C002 1, M060261C001 3, M060261C005 1, M060262C003 1, M060262C004 1, M060262C005 1, M060262C006 1, M060262C007 1, M060262C001 3, M060262C009 1, M018857C015 1, M060262C010 1, M018856C009 1, M060261C009 1, M060262C021 1, M018857C034 2, M060262C012 1, M060262C013 1, M060262C016 1, M018856C012 1, M060261C010 1, M060261C011 1, M056047C001 3, M056047C003 1, M056047C002 1, M060261C012 1, M060262C019 1, M060262C020 1, M060261C013 1); MiniMed 780G with the Guardian 4 Sensor System User Guide(M036461C004 2, M036461C001 2, M036461C007 1, M036461C003 1, M036461C002 1, M036461C006 1); MiniMed 780G with the Guardian Sensor (3) System User Guide(M010682C005 2, M010682C008 1); MiniMed 780G with the Guardian Sensor 3 System User Guide(M010682C002 2, M010682C004 1, M010682C007 1, M010682C003 1)
Product quantity
453,144
Voluntary mandated
Voluntary: Firm initiated
City
Northridge
State
CA
Country
United States
Event id
98186
Free daily digest + "My products" watchlist

Get the daily digest, and tell us what to watch. Everything is optional except your email.

Free allergen alerts:

Categories:

Privacy: we store only what you enter here, use it only to send recall alerts you asked for, never sell or share it, and you can unsubscribe anytime. Double opt-in: we will email you a confirmation link before sending anything; alerts start only after you confirm. Every email has unsubscribe and manage-watchlist links.