Medtronic MiniMed, Inc.: MiniMed 740G Insulin Pump (MMT-1811, MMT-1812, MMT-1861, MMT-1862)
Read the official notice at FDA (device) →
Get alerts for Medtronic MiniMed
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Medtronic MiniMed, Inc.
- Product
- MiniMed 740G Insulin Pump (MMT-1811, MMT-1812, MMT-1861, MMT-1862)
- Reason / hazard
- All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- Worldwide - US Nationwide distribution including in the states of OR, CA, AZ, NY,TX,FL, OH, MO, TN,NJ, NC, SC, IL, VA, WA, DC, GA, PA, MS,IA, MI, NH, OK, SD, MT, WI, WV, MA, LA, MN, RI, AR, ME, AL, IN, UT, KY, CO, ND, MD, CT, NM, KS, ID, NV, DE, NE, VT, AK, HI, WY, PR, VI and the countries of Taiwan, New Zealand, Australia, Bangladesh, Brunei Darussalam, Cambodia, Indonesia, Malaysia, Maldives, Philippines, Thailand, Vietnam, Singapore, Pakistan, Hong Kong, China, Argentina, Aruba, Barbados, Bermuda, Bolivia, Plurinational State Of, Brazil, Chile, Colombia, Costa Rica, Dominican Republic, Ecuador, El Salvador, Guatemala, Honduras, Mexico, Panama, Paraguay, Peru, Puerto Rico, Trinidad And Tobago, Uruguay, Venezuela, Canada, South Korea, India, Albania, Algeria, Andorra, Armenia, Austria, Azerbaijan, Bahrain, Belarus, Belgium, Bosnia And Herzegovina, Botswana, Bulgaria, Canary Islands, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Faroe Islands, Finland, France, French Guyana, French Polynes., Georgia, Germany, Gibraltar, Greece, Guadeloupe, Hungary, Iceland, Iran, Iraq, Ireland, Israel, Italy, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Kyrgyzstan, Latvia, Lebanon, Libya, Liechtenstein, Lithuania, Luxembourg, Macedonia, Malawi, Malta, Martinique, Mauritius, Mayotte, Moldavia, Montenegro, Morocco, Namibia, Netherlands, New Caledonia, Nigeria, Northern Ireland, Norway, Oman, Poland, Portugal, Qatar, Reunion, Romania, Russia, San Marino, Saudi Arabia, Serbia, Sierra Leone, Slovakia, Slovenia, South Africa, Spain, St.Pier,Miquel., Sudan, Sweden, Switzerland, Syrian Arab Republic, Tanzania, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Yemen, Zambia, Zimbabwe.
- Recall initiated
- 2026-02-13
- Code info
- Pump: UDI-DI(GTIN): User Guide Name(Part-Version): MiniMed 740G Insulin Pump: 763000504908, 763000504915, 763000505073, 763000484637, 763000505042, 763000504861, 763000629984, 763000504946, 763000504878, 763000559557, 763000504960, 763000505080, 763000505066, 763000629991, 763000505004, 763000504922, 763000504892, 763000505059, 763000504977, 763000505028, 763000505172, 763000505127, 763000505134, 763000505158, 763000505141, 763000505103, 763000505110, 763000504991, 763000505165, 763000504939, 763000504953, 763000504984, 763000505189, 763000851224, 763000851217, 763000769451, 763000769468: MINIMED 740G SYSTEM USER GUIDE mg/dL(M000149C002 1, M000149C004 1, M000149C005 1, M000150C005 1, M000150C002 1, M000150C021 2, M000149C016 1, M000149C021 1, M000150C004 1, M000150C025 1, M000150C030 1, M000149C014 1, M000149C006 1, M000150C003 1, M000150C015 1, M000150C028 1, M000150C018 1, M000150C022 1, M000150C007 1, M000149C017 1, M000149C008 1, M000150C008 1, M000150C032 1, M000150C029 1, M000150C012 1, M000150C016 1, M000150C010 1, M000149C011 1, M046726C001 3, M046726C002 3, M000149C020 1, M000149C025 1, M000150C024 1, M000150C023 1, M000150C013 1).
- Product quantity
- 24,946
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Northridge
- State
- CA
- Country
- United States
- Event id
- 98186
Free daily digest + "My products" watchlist