Cardinal Health 200, LLC: Cardinal Health Monoject Tuberculin Syringe, 1 mL Luer Lock, Catalog Number 1180100777
Read the official notice at FDA (device) →
Get alerts for Cardinal Health
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Cardinal Health 200, LLC
- Product
- Cardinal Health Monoject Tuberculin Syringe, 1 mL Luer Lock, Catalog Number 1180100777
- Reason / hazard
- The outer packaging and blister pack labeling identify the product as 1mL Luer Lock Tuberculin Syringe (Product Code 1180100777), however the syringes contained within the packages are U-100 Insulin syringes.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- Worldwide distribution - US Nationwide and the country of Canada.
- Recall initiated
- 2026-02-27
- Code info
- UDI/DI 10884521013414 (each), 20884521013411 (box), 30884521013418 (case), Lot Numbers: 531164X, 531162X
- Product quantity
- 198,900 units
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Waukegan
- State
- IL
- Country
- United States
- Event id
- 98534