Insulet Corporation: Omnipod 5 Pods. Model/Catalog number: PT-001662 ASM 5PK Pod STRL OPS G6G7

Medical DevicesSource: FDA (device) · Date: 2026-04-22

Read the official notice at FDA (device) →

Share on XFacebookRedditEmail

Get alerts for Insulet

Guides: Which agency handles which recall (FDA vs USDA vs CPSC vs NHTSA)

The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.

Company / brand
Insulet Corporation
Product
Omnipod 5 Pods. Model/Catalog number: PT-001662 ASM 5PK Pod STRL OPS G6G7
Reason / hazard
Due to a manufacturing defect, certain Omnipod 5 Pods from 49 lots have an internal soft cannula tear that results in insulin leaking into the Pod instead of being delivered to the user regardless of basal or bolus delivery. This defect results from damage to the unexposed portion of the soft cannula during manufacturing, which would result in a compromised fluid path. The primary failure mode is pump under-delivery due to loss of insulin to an internal leak; in some cases, the defect may also lead to pump shutoff and cessation of insulin delivery when leaked insulin contacts Pod circuity in a manner that results in an electrical short. Under-delivery of insulin (both basal and bolus insulin) or cessation of insulin put users at risk of hyperglycemia, and complications from acute and chronic hyperglycemia, including dehydration, blurry vision, nausea, vomiting, altered mental status, diabetic ketoacidosis (DKA), hyperglycemic hyperosmolar syndrome (HHS), or even death. Users may require hospitalization or medical intervention to treat severe adverse health consequences. Not all devices with the defect will issue an alarm or alert the user. If there is sufficient leakage of insulin to cause a short in the circuity, the Pod will issue a Hazard Alarm that stops all insulin delivery and alerts the user to replace their Pod. In addition, if a user s glucose is trending high and is not responding to insulin delivery, the system may reach the maximum amount of insulin microboluses allowed by the system and trigger the Automated Delivery Restriction (ADR) alert that tells users to check their blood glucose and take appropriate actions (i.e., ADR is a response to persistent hyperglycemia and maximum automated delivery constraints rather than a direct detection of the leak). The magnitude of under-delivery is unknown and based on multiple factors, including how much insulin is being delivered, whether an alarm and/or alert triggers, whether and when the user recognizes the device defect, the duration of Pod use, and the size of the tear.
Classification
Class I
Status
Ongoing
Distribution / sold at
Nationwide distribution.
Recall initiated
2026-03-12
Code info
Model/Catalog number: PT-001662 ASM 5PK Pod STRL OPS G6G7. UDI-DI: 10385083000527. Lot Codes: PH1U02252541, PH1U03282511, PH1U03282522, PH1U03312511, PH1U03312521, PH1U04012511, PH1U04012521, PH1U05052511, PH1U08162531, PH1U08182531, PH1U08182541, PH1U09242511, PH1U09242521, PH1U09242531, PH1U09252521, PH1U09252531, PH1U10152541, PH1U10162531, PH1U10162541, PH1U10172531, PH1U10172541, PH1U10182531, PH1U10182541, PH1U10202511, PH1U10202521, PH1U10202531, PH1U10202541, PH1U10212531, PH1U10212541, PH1U10222531, PH1U10222541, PH1U10232531, PH1U10232541, PH1U10242521, PH1U10242531, PH1U10242541.
Product quantity
1,240,115 units
Voluntary mandated
Voluntary: Firm initiated
City
Acton
State
MA
Country
United States
Event id
98586
Free daily digest + "My products" watchlist

Get the daily digest, and tell us what to watch. Everything is optional except your email.

Free allergen alerts:

Categories:

Privacy: we store only what you enter here, use it only to send recall alerts you asked for, never sell or share it, and you can unsubscribe anytime. Double opt-in: we will email you a confirmation link before sending anything; alerts start only after you confirm. Every email has unsubscribe and manage-watchlist links.