Ventec Life Systems, Inc.: VOCSN V+Pro package, REF: PRT-01198-000; VOCSN V+Pro unit, REF: PRT-01185-000
Read the official notice at FDA (device) →
Get alerts for Ventec Life Systems
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Ventec Life Systems, Inc.
- Product
- VOCSN V+Pro package, REF: PRT-01198-000; VOCSN V+Pro unit, REF: PRT-01185-000
- Reason / hazard
- Respiratory system intended to provide continuous/intermittent ventilator support, may not have been evaluated under all intended high-pressure conditions during production testing, so an oxygen leak condition could occur that may lead to reduced below intended levels of fraction of inspired oxygen delivery prior to and during ventilation and this oxygen-enriched environment may increase fire risk
- Classification
- Class I
- Status
- Ongoing
- Distribution / sold at
- US: OH, GA, CA, MO
- Recall initiated
- 2026-03-25
- Code info
- UDI-DI (REF/UDI-DI): PRT-01198-000/00855573007914, PRT-01185-000/00855573007877 Package/Device Serial Number (package serial number may include a "P" at the end): 152884, 152885, 152886, 152887, 152888, 152889, 152890, 152891, 152892, 152893, 152895, 152896, 152897, 152898, 152899, 152900, 152901, 152902, 152903, 152904, 152908, 152909, 152910, 152911, 152912, 152913, 152914, 152915, 152916, 152917, 152918, 152919, 152986, 152987, 155236, 156507, 156512, 156515, 156516, 156517, 156524, 156532
- Product quantity
- 42
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Bothell
- State
- WA
- Country
- United States
- Event id
- 98554