North American Rescue LLC.: AIDBAGs are first aid kits of convenience composed of individually labeled devices designed as specialty first aid kits designed for the app
Read the official notice at FDA (device) →
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- North American Rescue LLC.
- Product
- AIDBAGs are first aid kits of convenience composed of individually labeled devices designed as specialty first aid kits designed for the appropriate level provider to evaluate and treat patients at the point-of-injury / point-of-illness. The original manufacturer labeling, including single use indications, is maintained for each device, which may be used independently and at different times. These specific kits contain the McKesson TRUE Metrix PRO Professional Monitoring Blood Glucose System that were recalled as part of RES 98317 with (510(k): K140100).
- Reason / hazard
- Kits contain the McKesson TRUE Metrix PRO Professional Monitoring Blood Glucose System that has been recalled for an issue with the software design of the E-5 Error Code where the meter displays an E-5 error code for a very high blood glucose event (> 600 mg/dL) or when there is a test strip error, which can lead to operator confusion when evaluating patients and a delay in appropriate treatment. This is a downstream recall of Trividia Health Class I, Recall Event ID 98317.
- Classification
- Class I
- Status
- Ongoing
- Distribution / sold at
- US: CA, WA
- Recall initiated
- 2026-03-27
- Code info
- Kit REF/UDI-DI/Lot: 85-4661/00842209132089/85-4661122024; 85-4742/00842209133536/85-4742123024. Meter: UDI-DI: 20612479197217, all lots.
- Product quantity
- 21
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Greer
- State
- SC
- Country
- United States
- Event id
- 98697
Free daily digest + "My products" watchlist