AVID Medical, Inc.: Halyard CARDIAC CATH PACK SAN23CARDQ kit. Model Number: DRCC36.
Read the official notice at FDA (device) →
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- AVID Medical, Inc.
- Product
- Halyard CARDIAC CATH PACK SAN23CARDQ kit. Model Number: DRCC36.
- Reason / hazard
- Potential risk for Medline syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
- Classification
- Class I
- Status
- Ongoing
- Distribution / sold at
- US distribution to FL, IL, MO, MS, NC, NE, TX.
- Recall initiated
- 2026-03-13
- Code info
- Model Number: DRCC36. UDI-DI (Kit Lot Numbers): 10809160469088 (1659651).
- Product quantity
- 180 kits
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Toano
- State
- VA
- Country
- United States
- Event id
- 98686