⚠ RecallRadar📷 ScanGuidesOfficial U.S. recall notices, restated plainly

AVID Medical, Inc.: Halyard CATH LAB kit. Model Number: SACL75-01.

Medical DevicesSource: FDA (device) · Date: 2026-05-13

Read the official notice at FDA (device) →

Share on XFacebookRedditEmail

Get alerts for AVID Medical

Guides: Which agency handles which recall (FDA vs USDA vs CPSC vs NHTSA)

The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.

Company / brand
AVID Medical, Inc.
Product
Halyard CATH LAB kit. Model Number: SACL75-01.
Reason / hazard
Potential risk for Medline syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Classification
Class I
Status
Ongoing
Distribution / sold at
US distribution to FL, IL, MO, MS, NC, NE, TX.
Recall initiated
2026-03-13
Code info
Model Number: SACL75-01. UDI (Kit Lot Numbers): (01)20809160481377(17)280520(10)341730(30)004(241)UINR44-01 (341730), (01)20809160481377(17)280623(10)341742(30)004(241)SACL75-01 (341742), (01)20809160481377(17)280609(10)345855(30)004(241)SACL75-01 (345855), (01)20809160481377(17)280609(10)345863(30)004(241)SACL75-01 (345863).
Product quantity
172 kits
Voluntary mandated
Voluntary: Firm initiated
City
Toano
State
VA
Country
United States
Event id
98686
Free daily digest + "My products" watchlist

Get the daily digest, and tell us what to watch. Everything is optional except your email.

Free allergen alerts:

Categories:

Privacy: we store only what you enter here, use it only to send recall alerts you asked for, never sell or share it, and you can unsubscribe anytime. Double opt-in: we will email you a confirmation link before sending anything; alerts start only after you confirm. Every email has unsubscribe and manage-watchlist links.