Siemens Healthcare Diagnostics, Inc.: Dimension Creatinine Flex reagent cartridge. Material Number: 10872079.

Medical DevicesSource: FDA (device) · Date: 2026-05-06

Read the official notice at FDA (device) →

Share on XFacebookRedditEmail

Get alerts for Siemens Healthcare Diagnostics

Guides: Which agency handles which recall (FDA vs USDA vs CPSC vs NHTSA)

The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.

Company / brand
Siemens Healthcare Diagnostics, Inc.
Product
Dimension Creatinine Flex reagent cartridge. Material Number: 10872079.
Reason / hazard
Siemens Healthcare Diagnostics is recalling certain lots of Dimension Creatinine Flex reagent cartridge (Dimension CRE2) because a potential imprecision in the Dimension CRE2 quality control (QC) and patient sample results using lot numbers GA6307 and BA7005 on the Dimension System. This imprecision may lead to erroneously decreased or increased creatinine patient results
Classification
Class II
Status
Ongoing
Distribution / sold at
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Bahamas, Brazil, Cambodia, Canada, Chile, China, Columbia, Costa Rica, Ecuador, Egypt, Germany, Honduras, India, Indonesia, Korea, Kuwait, Lebanon, Malaysia, Mexico, Nicaragua, Pakistan, Paraguay, Philippines, Qatar, Saudi Arabia, Serbia, Slovenia, South Africa, Sri Lanka, Taiwan, Thailand, U.A.E., Ukraine, Uruguay, Vietnam, Yemen.
Recall initiated
2026-03-18
Code info
Material Number: 10872079. Model Number: DF33B. UDI: (01)00630414595009(10)GA6307(17)20261103, (01)00630414595009(10)BA7005(17)20270105. Lot Numbers: GA6307, BA7005.
Product quantity
18,080 units
Voluntary mandated
Voluntary: Firm initiated
City
Newark
State
DE
Country
United States
Event id
98666
Free daily digest + "My products" watchlist

Get the daily digest, and tell us what to watch. Everything is optional except your email.

Free allergen alerts:

Categories:

Privacy: we store only what you enter here, use it only to send recall alerts you asked for, never sell or share it, and you can unsubscribe anytime. Double opt-in: we will email you a confirmation link before sending anything; alerts start only after you confirm. Every email has unsubscribe and manage-watchlist links.