⚠ RecallRadar📷 ScanGuidesOfficial U.S. recall notices, restated plainly

Beta Bionics, Inc.: iLet Bionic Pancreas, REF: BB1001

Medical DevicesSource: FDA (device) · Date: 2026-05-06

Read the official notice at FDA (device) →

Share on XFacebookRedditEmail

Get alerts for Beta Bionics

Guides: Which agency handles which recall (FDA vs USDA vs CPSC vs NHTSA)

The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.

Company / brand
Beta Bionics, Inc.
Product
iLet Bionic Pancreas, REF: BB1001
Reason / hazard
insulin pump software versions 1.4.2, and 1.4.3 issue with Dexcom Continuous Glucose Monitoring (CGM) G7 sensor may result in delayed CGM reading then switch to non-CGM dosing, leading to only basal/requested meal doses provided, no correction doses/insulin reductions/suspensions for high/low glucose unless blood glucose entered for sensor life or replacement. Potential health risks to users include severe hyperglycemia with and without DKA and severe hypoglycemia.
Classification
Class II
Status
Ongoing
Distribution / sold at
US Nationwide distribution in the states of TX, PA, AL, MI, OH, FL, TN, GA.
Recall initiated
2026-03-12
Code info
Software versions 1.4.3, and 1.4.2
Product quantity
1,080
Voluntary mandated
Voluntary: Firm initiated
City
Irvine
State
CA
Country
United States
Event id
98475
Free daily digest + "My products" watchlist

Get the daily digest, and tell us what to watch. Everything is optional except your email.

Free allergen alerts:

Categories:

Privacy: we store only what you enter here, use it only to send recall alerts you asked for, never sell or share it, and you can unsubscribe anytime. Double opt-in: we will email you a confirmation link before sending anything; alerts start only after you confirm. Every email has unsubscribe and manage-watchlist links.